1.
Press
CHANNEL SELECT
key.
2.
Press
PROGRAM
soft key.
3.
Set key to program position or if Authorization Code is
enabled, enter
4
-digit code.
4.
Press
CHANGE MODE
soft key.
3-22
Programming
Alaris
®
System Directions for Use
Patient Controlled Analgesia Module Section
Changing Programming Parameters During an Infusion
A
25
mg/4 h
CONT
DOSE
[Conc]: 1 mg/mL
MAX
LIMIT
2
mg/h
CHANGE
MODE
Guardrails Drug Setup
IMS Pre-fill 30mL
20
minutes
LOCKOUT
INTERVAL
2
mg
PCA
DOSE
>Press CONFIRM
CONFIRM
Infusion Modes
(Continued)
Stopping a Loading, PCA or Bolus Dose
1.
Press
CHANNEL SELECT
key.
2.
Press
STOP LOAD, STOP PCA
or
STOP BOLUS
soft key as
NOTE:
Available soft key and stop confirmation screen are
dependent on the type of dose currently infusing and current
infusion mode.
>Press START to Close
Summary
[Conc]: 1 mg/mL
A
START
STOP
LOAD
1
mg
Loading
Dose:
Summary
Morphine
3.
To stop dose and resume current program, press
Yes
soft
key.
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......