Programming
2-49
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
NOTES:
Do not enter a patient weight if weight is not used in the
calculation.
Pump Module: At rates less than
10
mL/h, the rate is
displayed to two decimal places, and the
VTBI
can be entered
and is displayed to two decimal places.
Pump Module: In the Drug Calculation mode, the system
infuses at the calculated rate rounded to the nearest one-
hundredth of a mL per hour (as displayed on the programming
screen). The rate shown in the Rate Display will be rounded
to the nearest one-tenth of a mL per hour.
A
A
A
RATE
>Press START
VTBI
CONTINUOUS INFUSION
DOSE
5
mcg/kg/min
[Conc]: 5000 mcg/mL
4.2
mL/h
SETUP BOLUS START
PAUSE
Drug Calculation
50
mL
Pausing, Restarting, Restoring Infusion
Reference "Primary Infusion - With Guardrails
Protection", "Continuous Infusion" procedure.
Changing Rate or VTBI During Infusion
Reference "Primary Infusion - With Guardrails
Protection", "Continuous Infusion" procedure.
Infusion - NO
Guardrails
®
Suite MX Protection
(Continued)
Continuous Infusion - Drug Calculation
(Continued)
15. Verify parameters are correct and press
START
soft key.
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......