Programming
2-43
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Pausing, Restarting, Restoring Infusion
Reference "Primary Infusion - With Guardrails
Protection", "Continuous Infusion" procedure.
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
[Conc]: 342 mg/mL
4
mL/h
6:00
hh:mm
RATE =
VTBI
DURATION
RATE
VOLUME
A
>Verify Secondary Clamp
Open, Then Press START
DOSE =
SETUP
1200
mg/m
2
START
PAUSE
ALL
Guardrails Drug Setup
Methotrexate
SECONDARY
Infusion
(Continued)
7.
If an optional hospital-defined and editable starting value
for intermittent duration is not already entered, enter
duration or rate, as follows:
•
To enter duration, press
DURATION
soft key and use
numeric data entry keys (rate value is calculated and
displayed).
• To enter rate, press
RATE VOLUME
soft key and use
numeric data entry keys.
Changing Rate or VTBI During Infusion
Reference "Primary Infusion - With Guardrails
Protection", "Continuous Infusion" procedure.
Secondary Infusion
- With Guardrails
®
Suite MX Protection
(
Pump Module
)
(Continued)
8.
Verify parameters are correct and press
START
soft key.
• If programmed duration is outside Soft Limit for that
care area, a prompt appears before programming can
continue. If
Yes
soft key is pressed, programming
continues; if
No
soft key is pressed, infusion needs to
be reprogrammed.
•
If programmed duration is outside Hard Limit for that
care area, a prompt appears before programming can
continue. Infusion needs to be reprogrammed.
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......