Occlusion
Pressure Disc Installed
Pressure Disc Removed
Increased back pressure sensed
while infusing. Infusion stops on
affected module.
Pressure sensing disc installed
during an infusion. Infusion stops
on affected module.
Pressure sensing disc removed.
Infusion stops on affected module.
Clear occlusion. Press
RESTART
key, or press
CHANNEL SELECT
key and then
START
soft key.
Press
CONFIRM
soft key and
RESTART
key.
Reinsert pressure sensing disc and
press
RESTART
key.
Pump Chamber Blocked
Restart Channel
Blocked pump chamber detected.
Door opened and closed during
¤
an infusion. Infusion stops on
affected module.
Module paused for
2
minutes.
¤
Open door and inspect pump
chamber. To open blockage, as
required, massage tubing. To
continue infusion, press
RESET
soft
key and then
RESTART
key.
Close door. Press
RESTART
key,
or press
CHANNEL SELECT
key
and then
START
soft key.
Press
RESTART
key, or press
CHANNEL SELECT
key and then
START
soft key.
2-112
Troubleshooting and Maintenance
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Alarm
Meaning
Response
Syringe Module
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Alarms, Errors, Messages
(Continued)
Alarms
(Continued)
Alarm
Meaning
Response
Pump Module
(Continued)
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......