Display references throughout this procedure are for
illustration purposes only.
Reference "General Information", "Features and Displays" and
PC Unit Section of this
DFU
for information on the following:
• Displays
• Operating Features, Controls, Indicators
Programming
Programming
5-5
1.
Perform following steps (reference PC Unit Section of this
, "General Setup and Operation", "Start-Up"):
a. Power on system.
b. Choose
Yes
or
No
to
New Patient?
c. Confirm current profile or select a new profile.
d. Enter patient identifier, if required.
2.
Connect Microstream
®
Disposable (reference "Getting
3.
Press
CHANNEL SELECT
key.
•
SENSOR WARMING
and then
SEARCHING
appear in
Channel Message display until
E
t
CO
2
and respiratory
rate readings stabilize (up to
60
seconds).
Monitoring Mode
CHANNEL
SELECT
MONITOR
CHANNEL
OFF
EtCO (mm Hg)
2
RR (breaths/min)
4.
To change settings, reference "Setting Alarm Limits"
procedure.
OR
-- Continued on Next Page --
Alaris
®
System Directions for Use
EtCO
2
Module Section
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......