1-16
General Setup and Operation
Alaris
®
System Directions for Use
PC Unit Section
System Options
(Continued)
1.
Press
OPTIONS
key.
2.
Press
Power Down All Channels
soft key.
Power Down All Channels
System Options
Yes
No
Power Down
All Channels?
>Press Yes or No
S
y
stem Options 1 of 2
>Select an Option or
EXIT
PAGE
DOWN
EXIT
Displa
y
Contrast
Time of Da
y
Power Down All Channels
Anesthesia Mode
Patient ID
3.
Press
Yes
soft key.
• During power off sequence, Main Display flashes
POWERING DOWN
.
Anesthesia Mode
When the Anesthesia Mode is enabled while a module is
paused, the module remains in an indefinite pause until
restarted.
When Anesthesia Mode is enabled:
• All limits are set to
Soft
.
• Dose checking mode is set to
Smart
.
• Key-press audio is turned off.
• Tamper Resist Mode (panel locked) is not available.
• All Guardrails
®
Drug entries are available for selection.
When the Alaris
®
System is set up
for use in Anesthesia Mode, it is
important to
select the profile
that
corresponds with the care area the
patient will be taken to when the
Anesthesia Mode is discontinued.
This ensures that the Alaris
®
System
will be in the correct profile following
the use of the Anesthesia Mode.
CAUTION
-- Continued Next Page --
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......