General Information
4-27
Alaris
®
System Directions for Use
SpO
2
Modules Section
Specifications
(Continued)
Specifications and Symbols
(Continued)
8210 Series
Accuracy and Motion Tolerance:
Motion
No Motion
Pulse Rate:
20
-
250
bpm
normal physiologic range
20
-
250
±3
digits
(
55
-
125
bpm)
±5
digits
±3
digits
Functional Saturation:
70
-
100%
70
-
70
-
±2
digits
Adults, Neonates
Adults,
±2
digits
±3
digits
Neonates,
±3
digits
Display Update Period:
2.25
seconds
Measurement Range:
Perfusion
0.03
-
20%
Pulse Rate
20
-
250
bpm
S
p
O
2
1
-
100%
Pulse Amplitude Display:
Visual indicators for pulse signals represent proportional pulse amplitude
strength.
Sensor:
Emitted light wavelength range is within
500
-
1000
nm. Output power
does not exceed
15
mw.
(8210)
8220 Series
Accuracy and Motion Tolerance:
Resolution
Pulse Rate:
25
-
240
bpm
25
-
240
bpm
25
-
240
bpm
1
bpm
±3
digits
±5
digits
±3
digits
Adults, Pediatrics,
Adults, Pediatrics,
Adults, Pediatrics,
Neonates
Neonates
Neonates
Saturation:
70
-
100%
70
-
100%
70
-
100%
1% S
p
O
2
Adults, Pediatrics:
±3
digits
Adults, Pediatrics:
±2
digits;
Adults, Pediatrics,
±2
digits;
Neonates:
±3
digits
Neonates
Neonates:
±3
digits
Display Update Period:
Approximately
1
second.
(8220)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......