2-28
Programming
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
e. Press
BOLUS
soft key.
f. To enter bolus dose, use numeric data entry keys.
5.
Press
DURATION
soft key.
6.
To enter bolus duration, use numeric data entry keys.
OR
To deliver bolus dose at maximum safe rate possible for
selected drug and setup, and automatically calculate
bolus duration, press
Rapid Bolus
soft key.
•
TOTAL DOSE
alternates with
INFUSE AT
rate.
.09
mg/kg
A
A
A
DOSE
PATIENT
WEIGHT
BOLUS DOSE
DURATION
SETUP
>Enter Duration
70
kg
Rapid
Bolus
_ _
min
(420 mL/h)
TOTAL DOSE = 6.3 mg
BOLUS VTBI = 6.3 mL
Guardrails Drug Setup
Alteplase
CONT-
INUOUS
OR
A
A
A
CONTINUOUS INFUSION
Drug Calculation
Heparin
A
Dose will recalculate
based on new weight.
Adjust Dose or Rate
if required.
Accept weight change?
>Select Yes or No
Yes
No
Guardrails Drug Setup
Alteplase
d. Press
NEXT
soft key.
• If a continuous infusion is running, a prompt to
confirm weight change appears.
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
• When continuous dose is weight-based: (Continued)
Bolus Dose
(Continued)
Primary Infusion
- With Guardrails
®
Suite MX Protection
(Continued)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......