Alaris
®
System Directions for Use
Patient Controlled Analgesia Module Section
Programming
3-31
Infusion Modes
(Continued)
Stopping Infusion
Press and hold
CHANNEL OFF
key until a beep is heard,
approximately
1.5
NOTE:
If no other channel is active, the system powers down when
the
CHANNEL OFF
key is released.
Selecting Pressure Limit
1.
Press
CHANNEL SELECT
key.
2.
Press
OPTIONS
key.
3. Press
Pressure Limit
soft key.
A
>Select an Option or
EXIT
EXIT
Pressure Limit - High
Channel Options
2 of 2
Prime Set with S
y
ringe
PAGE
UP
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
4.
To select a pressure limit, press appropriate soft key.
5.
Press
CONFIRM
soft key.
A
Pressure Limit Selection
CONFIRM
Low
High
Med
NOTE:
Option to change pressure limit can be selected:
• after drug is selected, and before infusion mode is
selected and infusion starts, or
• after infusion starts.
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......