7.
To enter loading dose, press
LOAD DOSE
soft key, press
Yes
soft key and use numeric data entry keys.
8.
Verify parameters are correct and press
CONFIRM
soft
key.
• If programmed parameters are outside soft limit for that
care area, a prompt appears before programming can
continue. If
Yes
soft key is pressed, programming
continues; if
No
soft key is pressed, infusion needs to
be reprogrammed.
• If programmed parameters are outside hard limit for
that care area, a prompt appears before programming
can continue. Infusion needs to be reprogrammed.
9.
Close and lock security door.
10. Verify parameters on second nurse summary screen are
correct and press
START
soft key.
• During
PCA
dose + continuous infusion:
i
Green Infusing Status Indicator illuminates.
i
DELIVERING PCA
scrolls in Channel Message
Display when initiated. Continuous and
PCA
drug
name scrolls in Channel Message Display between
PCA
doses.
i
Volume infused for continuous dose is displayed in
mL/h in Rate Display.
i
Main Display alternates between volume remaining
and infusion mode with
PCA
drug name.
i
When
PCA
dose is complete,
PCA COMPLETE
scrolls in Channel Message Display and resumes
continuous dose.
3-18
Programming
Alaris
®
System Directions for Use
Patient Controlled Analgesia Module Section
Infusion Modes
(Continued)
Setting Up PCA Dose + Continuous Infusion
(Continued)
A
[Conc]: 1 mg/mL
10
minutes
MAX
LIMIT
2
mg/h
1
mg
CONFIRM
_
20
mg/4 h
PCA
DOSE
CONT
DOSE
LOADING
DOSE
None
CHANGE
MODE
Guardrails Drug Setup
IMS Pre-fill 30mL
LOCKOUT
INTERVAL
>Close and Lock Door Then
Press CONFIRM
A
[Conc]: 1 mg/mL
>Press START
Second Nurse Summar
y
Morphine
START
PAUSE PRO-
GRAM
10
minutes
Lockout
Interval:
Max Limit:
2
mg/h
1
mg
Cont. Dose:
PCA Dose:
8
mg/4 h
NOTES:
Time (in hours) associated with
Max Limit
is automatically
entered based on setup in system configuration.
Loading dose is included in
VTBI
but is not included in
Max
Limit
.
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......