Alaris
®
System Directions for Use
Patient Controlled Analgesia Module Section
Getting Started
3-5
Preparing and Loading Syringe and Administration Set
(Continued)
Loading Syringe and Administration Set
(Continued)
2.
Raise drive head to its fully extended position.
a. Twist gripper control clockwise and hold in position.
b. While holding gripper control in open position, raise
drive head to full extension.
c. Gently release gripper control.
Drive Head Fully Extended
Plunger Grippers Open
Gripper Control / Drive
Head Release in Open
Position
3.
Insert syringe (from front of instrument) by sliding flat
edge of syringe barrel flange between barrel flange
grippers.
4.
Lock syringe in place.
a. Pull syringe barrel clamp out and hold.
b. Rotate clamp to right (clockwise or counter clockwise)
until it lines up with syringe.
c. Gently release clamp against syringe.
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
PAUSE
RATE (mL/h)
20
25
30
1
oz
oz
60
35
40
45
50
1½
55
2
Syringe Barrel Clamp Closed
5
10
15
20
25
30
35
40
45
50
55
½
1
oz
1 ½
2
oz
60
50
55
2
oz
60
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......