Time to Alarm, Maximum
:
General Information
3-53
Alaris
®
System Directions for Use
Patient Controlled Analgesia Module Section
Rate (mL/h)
High
Low
1
120
minutes
37
minutes
5
30
minutes
7
minutes
Specifications and Symbols
(Continued)
Specifications
(Continued)
Maximum Time to Alarm specifications are based on
Alaris
®
Medical Systems
’
standard operating conditions:
Atmospheric Pressure:
645
-
795
mmHg
Back Pressure:
0
mmHg before producing occlusion
Humidity:
20
-
90%
Temperature:
68 ±4° F
Weight
:
5.5
lbs
Symbols
Reference the PC Unit Section of this
DFU
for system symbols.
Type
CF
, defibrillation-proof (PCA Module)
Type
BF
, defibrillation-proof (Dose Request Cord)
Manufacturer.
Single-Use. Do not re-use.
Product contains a particular element; such as,
=
DEHP
in fluid pathway.
Product
DOES NOT
contain a particular element; such as,
= administration set is
latex-free.
Approximate administration set priming volume.
LATEX
DEHP
»
XX ml
2
Single-Use
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......