Programming
2-29
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
5
mcg/kg
A
A
A
DOSE
PATIENT
WEIGHT
BOLUS DOSE
DURATION
[Conc]: 800 mcg/mL
SETUP
>Press Start to Begin
Infusing Bolus Dose
71
kg
<
1
min
START
PAUSE
CONT-
INUOUS
Guardrails Drug Librar
y
Dopamine
BOLUS VTBI = 50.0 mL
A
A
A
DOSE
BOLUS DOSE
DURATION
[Conc]: 1 mg/mL
SETUP
>Press START to Begin
Infusing Bolus Dose
1
min
START
PAUSE
CONT-
INUOUS
Guardrails Drug Setup
Alteplase
PATIENT
WEIGHT
INFUSE AT: 0.09 mg/kg/min
0.9
mg/kg
70
kg
Rapid
Bolus
(420 mL/h)
Bolus Dose
(Continued)
Stopping Bolus Dose
The display examples in this procedure represent stopping a
bolus dose which was programmed using the Drug Library.
Even where the displays are different when stopping a bolus
dose which was programmed using a non-library drug, the
procedure is the same.
1.
Press
CHANNEL SELECT
key.
2.
Press
STOP BOLUS
soft key.
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
7.
Verify parameters are correct and press
START
soft key.
• If programmed bolus dose and/or bolus dose duration
is outside Soft Limit for that care area, a prompt
appears before programming can continue. If
Yes
soft
key is pressed, programming continues; if
No
soft key
is pressed, infusion needs to be reprogrammed.
•
If programmed bolus dose and/or bolus dose duration
is outside Hard Limit for that care area, a prompt
appears before programming can continue. Infusion
needs to be reprogrammed.
•
If a bolus dose outside of Soft Limits has been entered
and verified as correct, Message Display also shows
either "
LLL
" for a low dose or "
↑↑↑
" for a high dose.
A
A
A
Dose
Remaining:
Time Left:
1
min
STOP
BOLUS
START
PAUSE
Pt. Weight:
70
kg
BOLUS DOSE
TOTAL DOSE = 6.3 mg
>Press START to Continue
Infusing Bolus Dose
BOLUS VTBI = 6.3 mL
Guardrails Drug Setup
Alteplase
0.9
mg/kg
Primary Infusion
- With Guardrails
®
Suite MX Protection
(Continued)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......