General Information
3-43
Alaris
®
System Directions for Use
Patient Controlled Analgesia Module Section
Compatible Syringes
Features and Displays
Features and Definitions
Reference the PC Unit Section of this
DFU
for system features and definitions.
Auto Pressure Limit Adjustment
When a bolus is delivered, pressure alarm limits are temporarily
raised to maximum limit.
Auto Syringe Identification
System automatically detects syringe size and narrows down
syringe selection list.
Bolus Delivery Rate
Rate at which
PCA
, bolus and loading doses (boluses) will be
infused.
Bolus Dose
Allows an additional amount of medication to be programmed
once
PCA
infusion has begun. Current
PCA
infusion resumes
following delivery of a bolus dose.
Continuous Dose
Basal rate dose.
The PCA Module is calibrated and labeled for use with the
following single-use disposable luer-lock syringes. Use only
the syringe size and type specified on the Main Display. The
full list of permitted syringe models is dependent on the PCA
Module’s software version.
20cc
30cc
35cc
50cc
60cc
Manufacturer
B-D Plastipak
IMS Pump Jet
Monoject
Terumo
NOTES:
Syringe variability may impact occlusion pressure sensing.
The variability may reduce the device’s time to alarm and/or
may require that a higher alarm pressure limit be
programmed.
Prefilled Morphine Sulfate
1
mg/ml.
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......