2-114
Troubleshooting and Maintenance
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Alarms, Errors, Messages
(Continued)
Alarms
(Continued)
Check Syringe
Drive Not Engaged
Plunger grippers opened during
¤
infusion and then closed. Infusion
stops on affected module.
Syringe barrel clamp opened
¤
during infusion and then closed.
Infusion stops on affected module.
Syringe plunger not captured
¤
while in idle state. System alarms
after
30
seconds, to indicate
potential siphoning condition.
Drive system disengaged during
operation.
Securely lock plunger grippers,
press
CHANNEL SELECT
key, and
reselect syringe.
Securely lock syringe barrel clamp
and press
RESTART
key.
Check for potential siphoning.
Ensure administration set clamp
(roller/slide) is in closed position.
Securely lock plunger grippers over
syringe plunger.
Open and close plunger grippers
and syringe barrel clamp. Ensure
syringe is properly installed.
Alarm
Meaning
Response
A
A
A
CONTINUOUS INFUSION
Drug Calculation
Heparin
A
S
y
ringe Installation
>Press CONFIRM
CONFIRM
plunger
flange
clamp
lever
• When problem is corrected, press
CONFIRM
soft key.
Syringe Adjustment Alarms
(Continued)
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......