• apnea
• very shallow breathing
• overmedication
• displaced cannula
No Breath Detected
• medication effect
• low tidal volume
Partial Airway Obstruction
• relaxation of upper airway
• head position
Hypoventilation with Shallow
Breathing
Waveform
Possible Causes
Waveform Analysis
(Continued)
Principle of Operation
The
E
t
CO
2
Module uses Oridion’s patented Microstream
®
nondispersive infrared (
NDIR
) spectroscopy to continuously
measure the amount of
CO
2
during every breath, the amount
of
CO
2
present at the end of exhalation (
E
t
CO
2
) and during
inhalation (
F
i
CO
2
), and the Respiratory Rate. The
E
t
CO
2
Module is a side stream capnograph.
The Microstream
®
Disposables deliver a sample of the inhaled
and exhaled gases from the ventilator disposable or directly
from the patient (via an oral/nasal cannula) into the monitor for
CO
2
measurement. Moisture and patient secretions are
extracted from the sample by the Microstream
®
inline filter
while maintaining the shape of the
CO
2
waveform.
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......