Programming
2-31
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Bolus Dose
(Continued)
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
NOTES
:
If the Bolus Dose feature is enabled, the
BOLUS
soft key
appears in the Continuous Infusion screen and becomes
active when a
VTBI
is entered.
The bolus
VTBI
cannot exceed the programmed continuous
infusion
VTBI
.
Programming and starting a bolus dose deletes any
programmed delay.
If no continuous rate is entered, the infusion will end when the
bolus has been delivered. No
KVO
infusion will follow.
To see details during the bolus infusion, press the
CHANNEL
SELECT
key.
The Pump Module keypad is used in the illustration but the
key is the same for the Syringe Module.
Restoring Bolus Dose
(Continued)
c. Press
RESTORE
soft key.
d. Verify dosing parameters and press
START
soft key.
2.
Bolus dose completed - module turned off:
a. Press
CHANNEL SELECT
key.
b. Press
RESTORE
soft key.
c. Verify parameters and press
NEXT
soft key.
d. Verify infusion parameters and press
BOLUS
soft key.
e. Press
RESTORE
soft key.
f. Verify dosing parameters and press
START
soft key.
Primary Infusion
- With Guardrails
®
Suite MX Protection
(Continued)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......