Principle of Operation
(Continued)
5-26
General Information
Alaris
®
System Directions for Use
EtCO
2
Module Section
Microstream
®
Disposables are single-use, disposables which
must be changed with each use. The manufacturer’s sample
flow,
50
ml/min, does not affect the disposable’s life; however,
humidity and specific patient conditions may shorten the
effective life of the disposables. Microstream
®
Disposables
are rated for up to
24
hours and
72
hours use, depending on
the specific Microstream
®
Disposable.
The
E
t
CO
2
Module provides readings in compliance with
BTPS
(body temperature, pressure, saturation) standards.
There is
no affect on accuracy due to cyclic pressure up to
10
kPa.
NOTE:
BTPS
(body temperature, pressure, saturation assumed
37°C
,
47
mmHg) calculations are made according to:
PCO
2
=
FCO
2
x (Pb -
47
)
Where:
FCO
2
is fractional concentration of
CO
2
in dry gas and
FCO
2
=
% CO
2
/
100
.
Pb is ambient pressure.
PCO
2
is partial pressure of
CO
2
at
BTPS
.
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......