1.
Press
VOLUME DURATION
soft key.
2.
To enter
VTBI
(Syringe Module: instead of infusing
ALL
)
,
press
VTBI
soft key and use numeric data entry keys.
OR
Syringe Module: To deliver entire contents of syringe,
leave
VTBI
as
ALL
3.
To enter volume duration, press
DURATION
soft key and
use numeric data entry keys.
• Rate is automatically calculated.
Programming
2-37
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
IV Fluid Infusion
(Continued)
4.
Verify correct infusion parameter entry and press
START
soft key.
• If programmed
IV
Fluid is outside Soft Limit for that
care area, a prompt appears before programming can
continue. If
Yes
soft key is pressed, programming
continues; if
No
soft key is pressed, infusion needs to
be reprogrammed.
•
If programmed
IV
Fluid is outside Hard Limit for that
care area, a prompt appears before programming can
continue. Infusion needs to be reprogrammed.
RATE =
4.17
mL/h
VTBI
DURATION
>Select DURATION
RATE
VOLUME
A
TPN
Guardrails Fluid Setup
PRIMARY INFUSION
ALL
12 00
h
min
START
DELAY
OPTIONS
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Volume/Duration Infusion
Primary Infusion
- With Guardrails
®
Suite MX Protection
(Continued)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......