Infusion of Air, Means to
Protect Patient from:
Ultrasonic Air-in-Line Detection
Maximum single bolus size = selectable
50
,
75
or
250
microliters nominal (
500
microliters in Anesthesia Mode)
Infusion Pressure,
Maximum:
654
mmHg (Maximum Occlusion Alarm Threshold plus tolerance)
KVO (Keep Vein Open)
Rate:
Factory Default Setting is
1
mL/h if set rate is
1
mL/h or above; or set rate, if rate
is
0.9
mL/h or below.
KVO Selection Range:
KVO
rate can be set in System Configuration from
0.1
-
20
mL/h in
0.1
mL/h
increments.
Occlusion Alarm Thresholds:
Pump Mode:
525
mmHg at rates
≥
30
mL/h
Varying level based on rate and patient back–pressure at rates
<30
mL/h.
Selectable Mode:
User selected,
50
-
525
mmHg in
25
mmHg increments.
Operating Principle:
Positive displacement
Rate Accuracy:
Rate accuracy of Alaris
®
System is
±5%
at rates between
1
and
999
mL/h and
±5.5%
at rates
<1
mL/h,
95%
of the time with
95%
confidence, under conditions
listed below.
Infusion Rate Range:
0.1
-
999
mL/h
Ambient Temperature:
68 ±4°F
(
20 ±2°C
)
Source Container Height:
20
inches above top of Pump Module
Test Solution:
Distilled Water
Distal Back pressure:
0
mmHg (
0
kPa)
Needle:
18
gauge
Administration Set Model
2210
2-96
General Information
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Variations of head height, back pressure or any combination of these
may affect rate accuracy
. Factors that can influence head height and
back pressure are: Administration set configuration,
IV
solution viscosity
and IV solution temperature. Back pressure may also be affected by type
of catheter. Reference "Trumpet and Start-Up Curves" for data on how
these factors influence rate accuracy.
WARNING
Specifications
(Continued)
Pump Module
(Continued)
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......