In this instrument, as with all infusion systems, the action of
the pumping mechanism and variations in individual
administration sets cause short-term fluctuations in rate
accuracy. The following graphs show typical performance of
the system, as follows:
• Accuracy during various time periods over which fluid
delivery is measured (trumpet curves).
• Delay in onset of fluid flow when infusion commences
(start-up curves).
Trumpet curves are named for their characteristic shape.
They display discrete accuracy data averaged over particular
time periods or "observation windows", not continuous data
versus operating time.
Over long observation windows, short-term fluctuations have
little effect on accuracy, as represented by the flat part of the
curve. As the observation window is reduced, short-term
fluctuations have greater effect, as represented by the
"mouth" of the trumpet. Knowledge of system accuracy over
various observation windows may be of interest when certain
drugs are being administered.
Because the clinical impact of short-term fluctuations on rate
accuracy depends on the half-life of the drug being infused
and on the degree of intravascular integration, the clinical
effect cannot be determined from the trumpet curves alone.
Knowledge of the start-up characteristics should also be
considered.
The start-up curves represent continuous flow rate versus
operating time for 2 hours from the start of the infusion. They
exhibit the delay in onset of delivery due to mechanical
compliance and provide a visual representation of uniformity.
Trumpet curves are derived from the second hour of this data.
Trumpet and Start-Up Curves
Introduction
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Pump and Syringe Modules
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......