5.
To enter loading dose, press
LOAD DOSE
soft key, press
Yes
soft key and use numeric data entry keys.
6.
Verify parameters are correct and press
CONFIRM
soft
key.
• If programmed parameters are outside soft limit for that
care area, a prompt appears before programming can
continue. If
Yes
soft key is pressed, programming
continues; if
No
soft key is pressed, infusion needs to
be reprogrammed.
• If programmed parameters are outside hard limit for
that care area, a prompt appears before programming
can continue. Infusion needs to be reprogrammed.
7.
Close and lock security door.
8.
Verify programming parameters are correct and press
START
soft key.
• Green Infusing Status Indicator illuminates.
• Infusion mode and drug name scroll in Channel
Message Display. If a loading dose has been entered,
DELIVERING LOAD
scrolls
.
• Volume infused in mL/h in Rate Display.
• Main Display alternates between volume remaining
and infusion mode with drug name.
3-16
Programming
Alaris
®
System Directions for Use
Patient Controlled Analgesia Module Section
Infusion Modes
(Continued)
Setting Up Continuous Infusion Only
(Continued)
[Conc]: 1 mg/mL
30
mg/4 h
Continuous
Morphine
A
>Close and Lock Door Then
Press CONFIRM
2
mg/h
CONFIRM
MAX
LIMIT
CONT
DOSE
LOADING
DOSE
_ _ 1
mg
CHANGE
MODE
A
[Conc]: 1 mg/mL
>Press START
Second Nurse Summary
Morphine
1
mg
START
PAUSE
PRO-
GRAM
Loading
Dose:
Max Limit:
2
mg/h
Cont. Dose:
30
mg/4 h
NOTES:
Time (in hours) associated with
Max Limit
is automatically
entered based on setup in system configuration.
Loading dose is included in volume infused but is not included
in
Max Limit
.
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......