Module has been designed and manufactured to
exacting standards and should perform well within given
environmental and performance standards. There may be
certain conditions under which an inaccurate measurement or
the loss of respiratory rate signal may occur. Examples of
these conditions are as follows:
An inaccurate
E
t
CO
2
measurement may be caused by:
• Incorrect disposable application or use.
• Microstream
®
Disposable disconnected or not securely
connected to
E
t
CO
2
Module.
• Airway connection clogged, twisted, or leaking.
• Placement too close to electrosurgery equipment.
• Mechanically ventilated patient breathes spontaneously.
5-22
General Information
Alaris
®
System Directions for Use
EtCO
2
Module Section
Measurement Accuracy
• If uncertain about
measurement
accuracy
, assess patient’s
condition and vital signs by
alternate means, then ensure
E
t
CO
2
Module is functioning
correctly.
•
Leaks or internal venting
of
sampled gas may affect
accuracy.
WARNINGS
Specifications
(Continued)
Specifications and Symbols
(Continued)
Warm-Up Time:
30
seconds typical
Weight:
2.5
lbs (
0.91
kg)
Reference the PC Unit Section of this
DFU
for system symbols.
Symbols
Type BF defibrillation-proof equipment.
Gas inlet.
Gas outlet.
Silenced alarm.
-- Continued on Next Page --
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......