Regulations and Standards
A-7
Alaris
®
System Directions for Use
Appendix Section
Table 2
Electromagnetic Immunity
IEC 61000-4-2
Electrostatic
Discharge (
ESD
)
Floors should be wood, concrete, or ceramic tile.
If floors are covered with synthetic material,
relative humidity should be at least
30%
.
If connector testing exemption is used, following
ESD
sensitivity symbol appears adjacent to each
connector.
" - Do Not Touch"
Emissions Test
Electromagnetic Environment - Guidance
Compliance
(Continued)
±6 kV contact
±8 kV air
IEC 60601-1-2
Test Level
±8 kV contact
Compliance
Level
IEC 61000-4-4
Electrical Fast
Transient, Burst
(EFT)
Mains power quality should be that of a typical
commercial or hospital environment.
±2 kV for power
supply lines
±1 kV for
input/output lines
±2 kV for power
supply lines
±1 kV for
input/output lines
IEC 61000-4-5
Mains power quality should be that of a typical
commercial or hospital environment.
±1 kV differential
mode
±2 kV common
mode
±1 kV differential
mode
±2 kV common
mode
IEC 61000-4-8
Power Frequency
Magnetic Field
(50/60 Hz)
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
3 A/m
400 A/m 50 Hz
400 A/m 60 Hz
-- Continued Next Page --
Alaris
®
System (Continued)
Electromagnetic Environment
(Continued)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......