4.
To enter time of day infusion is to be initiated (up to
23
hours
59
minutes), use numeric data entry keys.
5.
Press
CONFIRM
soft key.
• Time infusion is scheduled to start appears on Main
Display.
• If a
Before
callback has not been scheduled
(reference "Scheduling a Callback" procedure),
infusion automatically initiates at end of delay period.
The
Delay until
option is used to program an infusion delay
for a minimum of
1
minute and up to
23
hours
59
minutes.
1.
Press
DELAY OPTIONS
soft key.
2.
Press
Delay until
soft key.
Programming
2-59
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Delaying Infusion
(Continued)
NOTE:
If the current time has been previously confirmed, the
Time of
Day
screen will not be displayed.
Delay Options
(Continued)
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Specifying by Time of Day
A
Delay Options
08:00
CALL
BACK
Pause
CONFIRM
EXIT
>Select a Delay Option
Delay for
Delay until
Delayed Options
A
Time of Day
Change
Time
Current time:
08:00
CONFIRM
>CONFIRM Time-of-Day
EXIT
3.
If
Current time
displayed is correct, press
CONFIRM
soft
key; otherwise, press
Change Time
and enter correct
time. (Reference "System Options", "Time of Day" in PC
Unit Section of this
DFU
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......