3-52
General Information
Alaris
®
System Directions for Use
Patient Controlled Analgesia Module Section
Specifications and Symbols
(Continued)
Specifications
(Continued)
Flow Rate Programming
:
The flow rate range is from 0.1 to 999 mL/h as follows:
Flow Rates (mL)
Selectable Increments (mL/h)
0.10
-
9.99
0.01
10
-
99.9
0.1
100
-
999
1.0
Rate Restriction by Syringe Size:
Syringe Size (mL)
Flow Rate Range (mL/h)
50
/
60
0.1
-
999
30
/
35
0.1
-
650
20
0.1
-
500
Fluid Ingress Protection
:
IPX1, Drip Proof
Loading Dose Range
:
Configured according to hospital best-practice guidelines.
Max Limit Range:
Configured according to hospital best-practice guidelines.
Occlusion Alarm Thresholds
: Three settings:
Low
Medium
High
Operating Principle
:
Positive displacement
PCA Dose Range
:
Configured according to hospital best-practice guidelines.
Rate Accuracy
:
±2%
of full-scale plunger travel (not including syringe variation)
Syringe size and running force, variations of back pressure, or any combination
of these may affect rate accuracy. Factors that can influence back pressure are:
administration set configuration,
IV
solution viscosity, and
IV
solution
temperature. Back pressure may also be affected by type of catheter.
Reference “Trumpet and Start-Up Curves” for data on how these factors
influence rate accuracy.
WARNING
Shock Protection
:
Type
CF
, Defibrillator Proof (PCA Module)
Type
BF
, Defibrillator Proof (Dose Request Cord)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......