2-100
General Information
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Time to Alarm, Maximum:
Volume to be Infused
Programming Increments:
Syringe Module
(Continued)
Specifications
(Continued)
Rate (mL/h)
1
5
No Disc
High Setting
120 minutes
30 minutes
With Disc Highest
(1000 mmHg) Setting
105 minutes
30 minutes
Pressure Limit
Increments (mL)
0.01
0.1
Range (mL)
0.1
-
9.99
10
-
60
Weight:
4.5 lbs
NOTES:
On a high setting, the actual occlusion pressure will vary based on the syringe size and manufacturer.
Flow rates as low as
0.01
mL/h are available only with
1
cc and
3
cc syringes. For larger syringes, lower
limit adjusts to
0.1
mL/h.
The Maximum Time to Alarm specifications are based on Cardinal Health’s standard operating conditions:
Atmospheric Pressure:
645
-
795
mmHg
Back Pressure:
0
mmHg before producing occlusion
Humidity:
20
-
90%
Temperature:
68 ±4°F
Syringe Type:
BD 50
/
60
cc
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......