2.
Press
SECONDARY
soft key.
Programming
2-41
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
Infusion
(Continued)
Guardrails Fluid Setup
0.9% NaCI
RATE
VTBI
>Press START
40
mL/h
1000
mL
DELAY
OPTIONS
SECOND-
ARY
START
A
PRIMARY INFUSION
VOLUME
DURATION
BASIC
SEC
A
PAGE
DOWN
A
>Select Drug
Amifostine
Bleom
y
cin
C
y
tarabine
Daunorbicin
K-O
F-J
P-T
U-Z
A-E
Guardrails Drugs
Peds Oncolog
y
Azathioprine
PRIMARY
A
>Press Yes or No
Methotrexate
_ _ _ mg/_ _ _mL was
selected.
Is this correct?
Yes
No
mg/m
2
BSA
DOSING
UNITS
Guardrails Drug Setup
Peds Oncolog
y
Secondary Infusion
- With Guardrails
®
Suite MX Protection
(
Pump Module
)
(Continued)
3.
Press soft key next to desired drug.
• If applicable, an optional hospital-defined therapy or
clinical indication for delivery of this infusion may
appear. Different limits can be defined for same drug
with different therapeutic indications.
• If applicable, a weight-based, nonweight-based, or
BSA
-based option for delivery of this infusion may
appear.
• If applicable, multiple concentration listings for delivery
of this infusion may appear.
4.
To continue programming, press
Yes
soft key.
OR
To change selection, press
No
soft key.
• If
Yes
was selected and facility has defined a Clinical
Advisory for that drug, a message appears. To
indicate information has been noted and continue
programming, press
CONFIRM
soft key.
• If
Yes
was selected to continue programming, drug
amount and diluent volume (if defined in Drug Library)
are automatically entered for selected drug.
• If selected drug had "_ _ / _ _ mL" concentration, drug
amount and diluent volume need to be entered.
-- Continued on Next Page --
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......