Alaris
®
System Directions for Use
Patient Controlled Analgesia Module Section
Programming
3-13
Setting Up PCA Dose Only
PCA Dose only
Loading Dose Only
A
Continuous Infusion
>Select an Option
Guardrails Drug Setup
Morphine
PCA
Dose + Continuous
INFUSION MODES
SETUP
DRUG
LIBRARY
1.
Perform steps in “Preparing Infusion”.
2.
Press
PCA Dose Only
soft key from Infusion Mode
screen.
Infusion Modes
(Continued)
3.
To enter
PCA
dose, use numeric data entry keys.
[Conc]: 1 mg/mL
None
_ _ _
minutes
Morphine
PCA Only
MAX
LIMIT
A
_ 20
mg
LOADING
DOSE
None
CHANGE
MODE
PCA
DOSE
LOCKOUT
INTERVAL
>Select LOCKOUT
INTERVAL
Guardrails Drug Setup
IMS Pre-fill 30mL
[Conc]: 1 mg/mL
10
minutes
LOCKOUT
INTERVAL
MAX
A
>Select Max Limit Option
20
mg
PCA
DOSE
LOADING
DOSE
None
CHANGE
MODE
> > > >
Yes
No
LIMIT
4.
To enter lockout interval, press
LOCKOUT INTERVAL
soft
key and use numeric data entry keys.
5.
To enter maximum limit, press
MAX LIMIT
soft key and
then
Yes
soft key.
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......