A bolus dose can be restored after it has completed, either
prior to or after the module has been turned off, as indicated in
the following procedures.
The display examples in this procedure represent restoring a
bolus dose which was programmed using the Drug Library.
Even where the displays are different when restoring a bolus
dose which was programmed using a non-library drug, the
procedure is the same.
1.
Bolus dose completed - module not turned off:
a. Press
CHANNEL SELECT
key.
b. Verify infusion parameters and press
BOLUS
soft key.
2-30
Programming
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Restoring Bolus Dose
Bolus Dose
(Continued)
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
4.
To stop continuous infusion, press and hold
CHANNEL
OFF
key until a beep is heard (approximately
1.5
3.
To stop bolus and start continuous infusion, press
Yes
soft key.
A
>Press Yes or No
Yes
No
Stop Bolus and Start
Continuous infusion?
Guardrails Drug Setup
Alteplase
Stopping Bolus Dose
(Continued)
Primary Infusion
- With Guardrails
®
Suite MX Protection
(Continued)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......