at least 9½"
Hanger
Primary Container
Secondary Container
2-40
Programming
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Setup
(Continued)
Secondary Infusion
- With Guardrails
®
Suite MX Protection
(
Pump Module
)
(Continued)
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
The following procedure should be used only when:
• drug to be infused is listed in Guardrails
®
Drug Library,
• primary infusion is running, and
• a check valve administration set is being used.
To program a primary infusion, reference “IV Fluid Infusion”
procedure. To program a basic infusion, reference ”Infusion -
Suite MX Protection” procedure.
1.
Press
CHANNEL SELECT
key.
Infusion
6.
Using hanger provided with secondary administration set,
lower primary fluid container until bottom of secondary
container is at least
9½
" above fluid level in primary
container.