1.
Perform steps in “Preparing Infusion”.
2.
Press
PCA DOSE + CONTINUOUS
soft key from Infusion
Mode screen.
Alaris
®
System Directions for Use
Patient Controlled Analgesia Module Section
Programming
3-17
Infusion Modes
(Continued)
Setting Up PCA Dose + Continuous Infusion
PCA Dose onl
y
Loading Dose Onl
y
A
Continuous Infusion
>Select an Option
Guardrails Drug Setup
Morphine
PCA Dose + Continuous
INFUSION MODES
SETUP
DRUG
LIBRARY
A
[Conc]: 1 mg/mL
1
mg
_ _ _
mg/h
PCA
DOSE
CONT
DOSE
MAX
LIMIT
_ _ _
minutes
LOCKOUT
INTERVAL
>Enter Lockout Interval
LOADING
DOSE
None
CHANGE
MODE
PCA + Continuous
Morphine
None
3.
To enter
PCA
dose, press
PCA DOSE
soft key and use
numeric data entry keys.
4.
To enter lockout interval, press
LOCKOUT INTERVAL
soft
key and use numeric data entry keys.
5.
To enter continuous dose, press
CONT DOSE
soft key,
and use numeric data entry keys.
6.
To enter maximum limit, press
MAX LIMIT
soft key, press
Yes
soft key and use numeric data entry keys.
[Conc]: 1 mg/mL
10
minutes
LOCKOUT
INTERVAL
A
2
mg/h
1
mg
PCA
DOSE
CONT
DOSE
LOADING
DOSE
None
CHANGE
MODE
PCA + Continuous
Morphine
MAX
LIMIT
>Select Max Limit Option
> > > >
Yes
No
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......