General Information
3-45
Alaris
®
System Directions for Use
Patient Controlled Analgesia Module Section
Max Dose Limit
Optional configuration that limits total amount of drug allowed to
be delivered to patient in a defined period (
1
,
2
or
4
hours).
• This setting should be configured in data set
before
Drug
Library is developed. Once drugs are in Profile
PCA
Drug
Library, Max Accumulated Dose Limit cannot be changed.
• This optional setting applies to
all
drug setups within Profile
PCA Drug Library.
Module Location Enforcement
Tamper resistant security feature that ensures PCA Module is in a
tamper evident position. When enabled, PCA Module must be
located to direct right of PC Unit to allow programming an
infusion.
Near End of Infusion (NEOI)
Allows an alert to be configured to sound anywhere between
5 – 25%
volume remaining.
NEOI Alert
Alert Time can be set to occur when
5 – 25%
of
VTBI
remains.
Occlusion Pressure
Downstream occlusion alarm threshold can be set to low,
medium, or high.
Operating Modes
Four operating modes are available:
•
PCA
only
• continuous infusion
•
PCA
+ continuous infusion
• loading dose only
All programming of infusions in each of
4
modes are completed
using Drug Library as defined by hospital-established best-
practice.
Features and Displays
(Continued)
Features and Definitions
(Continued)
(Max Accumulated Dose Limit)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......