General Information
2-103
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Effects of Pressure Variations
Under conditions of
+100
mmHg pressure, the Pump Module
typically exhibits a long-term accuracy offset of approximately
-0.7%
from mean values.
Under conditions of
+300
mmHg pressure, the Pump Module
typically exhibits a long-term accuracy offset of approximately
-4.2%
from mean values.
Under conditions of
-100
mmHg pressure, the Pump Module
typically exhibits a long-term accuracy offset of approximately
+4.4%
from mean values.
Resulting trumpet observation points typically track those of
accuracy; therefore, no significant change in short-term
variations result under these pressure conditions.
Effects of Negative Solution Container Heights
With a negative head height of
-0.5
meters, the Pump Module
typically exhibits a long-term accuracy offset of approximately
-3.1%
from mean values.
Resulting trumpet observation points typically track those of
accuracy; therefore, no significant change in short-term
variations result under negative head height conditions.
Trumpet and Start-Up Curves
(Continued)
Introduction
(Continued)
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
Pump Module
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Syringe Module
Trumpet and start-up curves have been provided for
1.0
mL/h
and
5.0
mL/h. Measurements for trumpet curve rates below
1.0
mL/h are not provided because of the difficulty in
measuring extremely small volumes over a large duration of
time. In this case, the linear relationship of the plunger
position and velocity to syringe volume and rate is verified,
and is a function of the accuracy of the design.
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......