3-10
Programming
Alaris
®
System Directions for Use
Patient Controlled Analgesia Module Section
Preparing Infusion
(Continued)
Programming an Infusion
1.
Perform steps in “Getting Started”, “Preparing Syringe
and Administration Set”.
2.
Perform following steps (reference PC Unit Section of this
, “General Setup and Operation”, “Startup”):
a. Power on system.
b. Choose
Yes
or
No
to
New Patient?
c. Select profile, if required.
d. Enter patient identifier, if required.
3.
Press
CHANNEL SELECT
key.
4.
Unlock security door or set key to
PROGRAM
position.
5.
Confirm time of day or change time if necessary.
A
Time of Day
Change
Time
Current time:
08:00
>CONFIRM Time-of-Day
CONFIRM
PCA
EXIT
Priming
(Continued)
4.
When priming is complete, release
PRIME
soft key.
5.
To return to main screen, press
EXIT
soft key.
•
Guardrails Drug Setup
screen displays.
6.
Select infusion mode
.
NOTE:
Fluid is delivered during priming only while the
PRIME
soft key
is pressed. Each press of the
PRIME
soft key delivers up to
2
mL of priming/fluid per continuous press. To deliver
additional amounts, press the
PRIME
soft key again.
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......