Principle of Operation
(Continued)
4-32
General Information
Alaris
®
System Directions for Use
SpO
2
Modules Section
When saturation is calculated from a blood gas partial
pressure of oxygen (
PO
2
), the calculated value may differ from
the
S
p
O
2
measurement of the
S
p
O
2
Module, this usually
occurs because the calculated saturation was not
appropriately corrected for the effects of variables that shift the
relationship between
PO
2
and p
H
, temperature, the partial
pressure of carbon dioxide (
PCO
2
),
2,3-DPG
, and fetal
hemoglobin. To compare functional saturation measurements
to those from an instrument that measures fractional
saturation, fractional measurements must be converted as
follows:
Oxyhemoglobin Dissociation Curve
pH
Temperature
PCO
2,3-DPG
Fetal Hb
2
¯
¯
¯
pH
Temperature
PCO
2,3-DPG
2
¯
¯
¯
100
50
0
50
100
PO (mmHg)
2
Saturation
(%)
¬
®
8210 Series
(Continued)
(8210)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......