Alaris
®
System Directions for Use
Patient Controlled Analgesia Module Section
Programming
3-29
A
[Conc]: 1 mg/mL
START
Second Nurse Summar
y
Morphine
PAUSE
PRO-
GRAM
20
minutes
Lock
out
Interval:
Max
Limit:
2
mg
PCA Dose:
25
mg/4 h
BOLUS
DOSE
>Press START
Infusion Modes
(Continued)
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
From
Second Nurse Summary
screen, press
PAUSE
soft
key.
•
PAUSE
scrolls in Channel Message Display.
•
PAUSED
appears on Main Display.
• Yellow Standby Status Indicator illuminates.
• After
2
minutes,
PAUSE-RESTART CHANNEL
visual and
audio prompts begin, and yellow Standby Status
Indicator flashes.
Pausing Infusion
(Continued)
2.
To reinitiate infusion:
• Press
RESTART
key.
OR
• Press
CHANNEL SELECT
key and then press
START
soft key on Main Display.
Changing Syringe and Restoring Infusion
1.
If syringe requires replacement:
a. Unlock security door.
b. Remove existing syringe and prepare new syringe
(reference “Getting Started”, “Preparing and Loading
Syringe and Administration Set”).
c. Load syringe and administration set (reference
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......