Alaris
®
System Directions for Use
Patient Controlled Analgesia Module Section
Programming
3-35
A
4
RR
LOW
PCA Pause Limits
DISABLE
ETCO2
DISABLE
SPO2
CONFIRM
88
%SPO2
LOW
97
5
SPO2/ETCO2
ALARM LIMITS
PCA
PAUSE LIMITS
>Press CONFIRM to Apply
Changes
Reviewing or Changing PCA Pause Alarm Limits
(Continued)
4.
Verify
PCA
pause limits as per facility protocol or
physician order.
A
88
>Enter a Pause Limit
Between 2 - 4
CONFIRM
_4
PCA Pause Limits
DISABLE
ETCO2
DISABLE
SPO2
%SPO2
LOW
RR
LOW
97
5
SPO2/ETCO2
ALARM LIMITS
PCA
PAUSE LIMITS
NOTE:
The acceptable range for PCA Pause Protocol is configurable
and defined by the hospital within the data set using the
Guardrails
®
Suite MX.
The
PCA PAUSE LIMITS
must be lower than the
SPO2
/
ETCO2
ALARM LIMITS
. A prompt is provided if the
PCA PAUSE LIMITS
must be modified.
5.
To change
PCA
pause limits, press soft key that
corresponds to alarm limit and enter a value within
acceptable range.
6.
Press
CONFIRM
soft key.
7.
Press
START
soft key.
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......