CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
2-24
Programming
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Pausing and Restarting Infusion
1.
Press
PAUSE
key.
•
PAUSE
scrolls in Message Display.
•
PAUSED
appears on Main Display.
• Yellow Standby Status Indicator illuminates.
• After
2
minutes,
PAUSE-RESTART CHANNEL
visual and
audio prompts begin, and yellow Standby Status
Indicator flashes.
A
>Press REPROGRAM
Reprogram
Dose exceeds
Guardrails hard
limit of .82 mg/kg/h.
Guardrails Drug Setup
Alteplase
•
If programmed continuous dose infusion is outside
Guardrails
®
Hard Limit ("Hard Limit") for that care area,
a prompt appears before programming can continue.
Infusion needs to be reprogrammed.
•
If a dose outside of Soft Limits has been entered and
verified as correct, Message Display also shows either
"
LLL
" for a low dose or "
↑↑↑
" for a high dose.
Continuous Infusion
(Continued)
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
A
>Press Yes or No
Yes
No
Dose exceeds
Guardrail limit of
.81 mg/kg/h.
Proceed?
Guardrails Drug Setup
Alteplase
• If programmed continuous dose infusion is outside
Guardrails
®
Soft Limit ("Soft Limit") for that care area, a
prompt appears before programming can continue. If
Yes
soft key is pressed, programming continues; if
No
soft key is pressed, infusion needs to be
reprogrammed.
Primary Infusion
- With Guardrails
®
Suite MX Protection
(Continued)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......