220
SAW sensor – Parameter
Type 8098 FLOWave L
15.12.6
Activating the monitoring of the differentiation factor
→
Before activating the monitoring of the DF, set the DF error and warning limits. See chapter 15.12.8
By default, the monitoring of the DF and the diagnostics are all disabled.
To activate the monitoring of the DF, do the following:
→
Go to the
CONFIGURATION
view.
→
SAW sensor
→
Confirm to access the
Parameter
view.
→
Add. meas. values
→
DF
→
Limits
→
Active
Yes
.
→
Save.
The monitoring of the DF is active and the device status will change depending on the limits that have
been set.
→
You can transmit the DF value with an analogue output to a PLC for example to identify the liquid flowing
through the pipe.
→
You can configure the behaviour of an analogue output depending on the status of the device. See
chapter 18.3.3
.
→
You can configure a digital output to switch every time a specific event is generated. See chapter 18.5.1
Configuring a digital output as an on/off output
.
→
To enable the monitoring, i.e. to be informed when the value of the DF is outside the normal range,
enable the diagnostics. See chapter 12.10 Activating the diagnostics function
15.12.7
Deactivating the monitoring of the differentiation factor
By default, the DF values are not monitored. If the monitoring of the DF is active, do the following to deac-
tivate it:
→
Go to the
CONFIGURATION
view.
→
SAW sensor
→
Confirm to access the
Parameter
view.
→
Add. meas. values
→
DF
→
Limits
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...