232
SAW sensor – Parameter
Type 8098 FLOWave L
15.13.8
Deactivating the monitoring of the acoustic transmission factor
By default, the acoustic transmission factor values are not monitored.
But if the monitoring of the acoustic transmission factor is active, do the following to deactivate it:
→
Go to the
CONFIGURATION
view.
→
SAW sensor
→
Confirm to access the
Parameter
view.
→
Add. meas. values
→
Acoustic transmission factor
→
Limits
→
Active
No
.
→
Save.
The monitoring of the acoustic transmission factor is inactive.
15.13.9
Resetting the default values of the error limits, the warning limits
and the hysteresis of the acoustic transmission factor
The default values of the error limits, the warning limits and the hysteresis of the acoustic transmission
factor are the following:
•
high error value: 195 %,
•
low error value: 5 %,
•
high warning value: 190 %,
•
low warning value: 10 %,
•
value of the hysteresis: 1 %.
To reset the default values of the error limits, the warning limits and the hysteresis of the acoustic trans-
mission factor, do the following:
→
Go to the
CONFIGURATION
view.
→
SAW sensor
→
Confirm to access the
Parameter
view.
→
Add. meas. values
→
Acoustic transmission factor
→
Limits
→
Reset to default
Confirm.
The limit values and the hysteresis value are reset.
→
Go back to the parent menu.
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...