247
SAW sensor – Parameter
Type 8098 FLOWave L
15.16
Setting the refresh time
15.16.1
Use case of the refresh time
The refresh time is the minimum time needed to update a measurement value. The refresh time has no
effect on the damping of the measured values.
The refresh time of the temperature values is a constant but the refresh time of the other measurement
values can be adapted to the process:
•
A very short refresh time is needed if the process requires quick measurement updates, for example for
very short dosings.
•
A long refresh time is sufficient if for example there are slow flow rate changes in the process.
15.16.2
Changing the refresh time
3 refresh times are available:
• a
Long
refresh time, i.e. the duration between 2 measurement updates is approximately 100 ms.
• a
Short
refresh time, i.e. the duration between 2 measurement updates is approximately 70 ms. Default setting.
• a
Very short
refresh time, i.e. the duration between 2 measurement updates is approximately 30 ms.
If the very short refresh time is set:
• The diagnostics event
Not totally filled
is not available.
•
The measurement deviation for a flow rate between 10 % of the full scale and the full scale is ±0.6 %.
•
The repeatability for a flow rate between 10 % of the full scale and the full scale is ±0.3 %.
If a digital output is configured as a pulse output, the following durations must be added to the last
received pulse:
•
50 ms, if the refresh time is set to
Very short
,
•
80 ms, if the refresh time is set to
Short
,
•
140 ms, if the refresh time is set to
Long
.
To change the refresh time, do the following:
→
Go to the
CONFIGURATION
view.
→
SAW sensor
→
Confirm to access the
Parameter
view.
→
Refresh time
.
→
Choose the refresh time.
→
Save.
The refresh time is changed.
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...