173
SAW sensor – Parameter
Type 8098 FLOWave L
15.6
Setting the parameters of the liquid temperature
15.6.1
Giving a user defined name to the measured liquid
temperature
The name is used to identify the process value in the user defined views and in all the menus where the
process value is displayed (for example in the
Outputs
menu).
By default, the name associated to the measured liquid temperature is
Temperature
.
To add a user defined name to the default name, do the following:
→
Go to the
CONFIGURATION
view.
→
SAW sensor
→
Confirm to access the
Parameter
view.
→
Stand. meas. values
→
Temperature
→
Value name
.
→
Enter the name by selecting and confirming each character. The name can have up to 19
characters.
→
OK
→
Save the name.
The name is changed.
15.6.2
Activating the damping of the liquid temperature values and
choosing a predefined damping level
The damping makes it possible to damp the fluctuations of the measured values of the liquid temperature:
• on the display,
• on the outputs. The damping of the liquid temperature comes in addition to the damping set for each
analog output (see chapter 18.3.2
The damping is not applied to the new measured value, if the 2 following conditions are met:
• a
Low
,
Medium
or
High
damping level is active,
•
and the variation between 2 values that are measured one after the other is higher than 20 °C.
The refresh time, set in chapter 15.16, has no effect on the damping of the measured values.
By default, the measured liquid temperature values are not damped.
→
As an alternative to the 3 predefined damping levels
Low
,
Medium
or
High
, you can set your own
damping parameters. See
.
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...