Message "An error occurred during communication"
Message
An error occurred during communication
Symbol displayed in the
information bar
Possible cause
The flat cable that connects the sensor to the transmitter may be damaged.
What to do?
→
Make sure the flat cable is correctly connected.
→
If the cable is damaged, send the device back to Bürkert.
24.9.20
Message "Max. flow rate"
Message
Max. flow rate
Symbol displayed in the
information bar
(symbol not linked to a user setting)
Possible cause
The maximum flow rate is measured in the pipe. The flow rate in the tube is
higher than 10 m/s, whatever the DN of the tube.
What to do?
→
Make sure the flow rate value is less than 10 m/s.
24.9.21
Message "Max temperature"
Message
Max temperature
Symbol displayed in the
information bar
(symbol not linked to a user setting)
Possible cause
The temperature in the tube is higher than 150 °C. The maximum liquid tem
-
perature is measured in the pipe.
This message does not depend on the liquid temperature limits set by the
user.
What to do?
→
Make sure the liquid temperature is in the permitted range.
24.9.22
Message "Totalizer 1 stopped" / "Totalizer 2 stopped" or "Mass
totalizer 1 stopped" / "Mass totalizer 2 stopped"
Message
Totalizer 1 stopped / Totalizer 2 stopped
or
Mass totalizer 1 stopped / Mass totalizer 2 stopped
Symbol displayed in the
information bar
-
Possible cause
The related totalizer / mass totalizer has been stopped by the user.
What to do?
→
If needed, start the totalizer
/ mass totalizer again.
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...