133
General settings – Diagnostics
Type 8098 FLOWave L
→
Transmit errors
→
Go back to the parent menu.
13.3.4
Reading out the maximum number of transmit errors since the
last power-up of the device
To read out the maximum number of transmit errors, do the following:
→
Go to the
CONFIGURATION
view.
→
General settings
→
Confirm to access the
Parameter
view.
→
Go to the
DIAGNOSTICS
view.
→
büS status
Transmit errors max.
→
Go back to the parent menu.
13.3.5
Resetting the 2 maximum error counters
To reset the 2 maximum error counters, do the following:
→
Go to the
CONFIGURATION
view.
→
General settings
→
Confirm to access the
Parameter
view.
→
Go to the
DIAGNOSTICS
view.
→
büS status
→
Reset error counter
→
Confirm.
The 2 maximum error counters are reset.
13.3.6
Reading out whether the measured process data (PDO,
process data object) is sent on büS or on the CANopen
fieldbus
To read out whether the measured process data (PDO, process data object) is sent on büS or on the
CANopen fieldbus, do the following:
→
Go to the
CONFIGURATION
view.
→
General settings
→
Confirm to access the
Parameter
view.
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...