340
Maintenance and troubleshooting
Type 8098 FLOWave L
24.11.14
Message "Density too low"
Message
Density too low
Symbol displayed in the
information bar
Possible cause
The value of the density is lower than the permitted minimum error value set
in chapter 15.9.7 Changing the error limits, the warning limits and the hys
teresis of the liquid density.
What to do?
As soon as the value of the density returns to within the permitted range,
the error is automatically reset.
Message
Density too low
Symbol displayed in the
information bar
Possible cause
The value of the density is lower than the permitted minimum warning value
set in chapter 15.9.7 Changing the error limits, the warning limits and the
hysteresis of the liquid density.
What to do?
As soon as the value of the density returns to within the permitted range,
the warning is automatically reset.
24.11.15
Message "Concentration 1 too high" / "Concentration 2 too high"
Message
Concentration 1 too high / Concentration 2 too high
Symbol displayed in the
information bar
Possible cause
The value of the liquid concentration is higher than the permitted maximum
error value set in chapter 15.7.8 Changing the error limits, the warning limits
and the hysteresis of the liquid concentration.
What to do?
As soon as the value of the liquid concentration returns to within the per-
mitted range, the error is automatically reset.
Message
Concentration 1 too high / Concentration 2 too high
Symbol displayed in the
information bar
Possible cause
The value of the liquid concentration is higher than the permitted maximum
warning value set in chapter 15.7.8 Changing the error limits, the warning
limits and the hysteresis of the liquid concentration.
What to do?
As soon as the value of the liquid concentration returns to within the per-
mitted range, the warning is automatically reset.
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...