213
SAW sensor – Parameter
Type 8098 FLOWave L
15.11.8
Disabling the user to start, stop or reset each mass totalizer
By default, the user is not allowed to start, to stop or to reset a mass totalizer.
If the Start/Stop/Reset of a mass totalizer is active, do the following to disable them:
→
Go to the
CONFIGURATION
view.
→
SAW sensor
→
Confirm to access the
Parameter
view.
→
Stand. meas. values
→
Mass totalizer 1
or
Mass totalizer 2
→
Start/Stop/Reset
→
Disabled
→
Save.
The user is not authorized to start or to stop or to reset a mass totalizer.
15.11.9
Starting each mass totalizer
If the Start/Stop/Reset of a mass totalizer is active, do the following to start the mass totalizer:
→
Go to the
CONFIGURATION
view.
→
SAW sensor
→
Confirm to access the
Parameter
view.
→
Stand. meas. values
→
Mass totalizer 1
or
Mass totalizer 2
→
Start/Stop
→
Started
→
Save.
The mass totalizer starts to count.
15.11.10
Stopping each mass totalizer
If the Start/Stop/Reset of a mass totalizer is active, do the following to stop the mass totalizer:
→
Go to the
CONFIGURATION
view.
→
SAW sensor
→
Confirm to access the
Parameter
view.
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...