SAW sensor – Parameter
Type 8098 FLOWave L
15.13.7
Activating the monitoring of the acoustic transmission factor
To be informed when the concentration of gas bubbles or solid particles changes in the liquid, monitor the
acoustic transmission factor.
A monitored value can be:
• in the normal operating range.
• in the warning range,
• in the error range.
You can set 4 limit values: 2 error limits and 2 warning limits.
→
To set the limit values, see chapter 15.13.6 Changing the error limits, the warning limits and the hys
teresis of the acoustic transmission factor.
15.4.5 explains how the device reacts when the monitored value enters in another range (for
example, from the normal range into the warning range). The reaction time depends on the hysteresis value
and if the monitored value increases or decreases.
By default, the monitoring of the acoustic transmission factor and the diagnostics are all disabled.
To activate the monitoring of the acoustic transmission factor, do the following:
→
Go to the
CONFIGURATION
view.
→
SAW sensor
→
Confirm to access the
Parameter
view.
→
Add. meas. values
→
Acoustic transmission factor
→
Limits
→
Active
→
Yes
.
→
Save.
The monitoring of the acoustic transmission factor is active and the device status will change depending
on the limits that have been set.
→
You can configure the behaviour of an analogue output depending on the status of the device. See
chapter 18.3.3
.
→
You can configure a digital output to switch every time a specific event is generated. See chapter 18.5.1
Configuring a digital output as an on/off output
.
→
To enable the monitoring, i.e. to be informed when the value of the acoustic transmission factor is
outside the normal range, enable the diagnostics. See chapter 12.10 Activating the diagnostics function
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...