SAW sensor – Maintenance
Type 8098 FLOWave L
→
Calibration
→
Add. meas. values
→
Concentration 1
or
Concentration 2
→
Offset
→
Teach-in by reference
The
Current settings
are displayed.
→
Start the calibration of the offset value.
→
After 30 s, enter the concentration of the reference liquid.
The
New settings
are displayed.
→
Save.
The offset value of the selected liquid concentration is calibrated.
If the calibration fails, a message is displayed. Refer to chapter 24.10 Messages due to calibration or
→
Acknowledge the message to go back to the parent menu.
17.22
Setting the slope value of the liquid concentration
To enter a slope value for liquid concentration 1 or liquid concentration 2, do the following:
→
Go to the
CONFIGURATION
view.
→
SAW sensor
→
Confirm to access the
Parameter
view.
→
Go to the
MAINTENANCE
view.
→
Calibration
→
Add. meas. values
→
Concentration 1
or
Concentration 2
→
Slope
→
Enter value
→
Set the slope value.
→
Save.
The slope value of the selected liquid concentration is set.
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...