120
General settings – Parameter
Type 8098 FLOWave L
12.11
Disabling all the diagnostics
By default, all the diagnostics events related to the process, the electronics or the sensor, the messages
related to the monitoring of the process values (e.g. the flow rate) and the messages related to problems on
the device and on büS are disabled.
If the diagnostics are active on the device, do the following to disable them:
→
Go to the
CONFIGURATION
view.
→
General settings
→
Confirm to access the
Parameter
view.
→
Diagnostics
Read the displayed message
→
Inactive
→
Save and restart the device.
All the diagnostics are disabled.
12.12
Configuration provider
If active on the device, the configuration provider manages the configuration data of the device modules
(for example, display module, Ethernet module,...). The configuration provider does not manage the con
-
figuration data of the transmitter board. The transmitter board directly saves its configuration data on the
configuration memory.
→
Make sure the configuration provider is active on the device. Refer to chapter 12.12.1
.
→
Make sure a configuration memory (SIM card) is inserted in its slot on the device. Refer to Fig. 50
or to
the parameter
Transferable memory status
described in chapter 13.2.9 Reading out the status of the
.
→
Make sure you use a configuration memory from Bürkert. Configuration memories can be bought from
your Bürkert branch office.
Upon the device startup, the device can have one of the following behaviours:
•
If the configuration memory is empty or contains data from a device with a previous software version, the
configuration memory is formatted and the current configuration data is saved on it.
•
If the configuration memory contains data which is compatible with the configuration provider, the serial
numbers of the modules with the same article number are compared:
− If the serial numbers are different, the configuration provider copies the configuration data from the
configuration memory to the device modules.
− If the serial numbers are the same, the configuration data are not copied.
− If the device is equipped with an additional module, its configuration data is saved on the configuration
memory.
•
The configuration provider automatically saves the configuration data of a module as soon as a module
setting has been changed.
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...